Impact of adding a filter for protection from toxic inhalational compounds to the ventilation circuit of mechanically ventilated patients

Abstract

Background

Standard-issue Chemical-Biological-Radio-Nuclear (CBRN) gasmasks, as used for protection from non-conventional warfare agents or toxic industrial compounds, cannot be used by ventilated patients, leaving them exposed to toxic agents inhaled via their ventilators. This study was conducted to determine the safety of a CBRN filter added to the patient circuit of a ventilator, as a method for affording inhalational protection to ventilated patients.

Methods

A Landrace pig was ventilated sequentially with 3 types of ventilators according to 17 different ventilation protocols, with and without a CBRN filters added in-line to the ventilation tubing for each protocol. For each protocol, physiological parameters, including oxygen saturation, inspired CO2, end tidal CO2, inspired oxygen, respiratory rate, and pulse rate, as well as airflow parameters including peak inspiratory pressure, positive end expiratory pressure and tidal volume were measured. The impact on the ventilator’s trigger/sensitivity function was evaluated in vitro using a Michigan test lung.

Results

On average, the addition of the CBRN filter resulted in a 16 ml (5 %) decrease (range 0–50 ml) in the tidal volume, a 1.7 cm H2O (10 %) decrease (range 1–3 cm H2O) in the peak inspiratory pressure, and a 0.1 cm H2O (3 %) decrease (range 0–1 cm H2O) in the positive end expiratory pressure delivered to the animal. Some ventilators compensated for these airflow changes while others did not, depending on the design of the ventilator’s pressure/flow sensing mechanism. Significant rebreathing occurred when the filter was positioned directly on the animal’s endotracheal tube, but not when positioned on the air outflow port of the ventilator. In vitro measurements showed that the addition of the CBRN filter added a mean pressure gradient of 0.45 cm H2O to the trigger/sensitivity function of the system.

Conclusions

In-line addition of a CBRN filter to ventilation tubing is a feasible strategy for affording inhalational protection to ventilated patients.

Background

Dispersion of volatile Toxic Industrial Compounds (TICs) presents a major threat to nearby residents. A scenario of volatile TIC dispersion may occur due to an industrial accident, or as a result of an act of terrorism [110]. Protection against TIC exposure can be afforded by wearing a standard issue gasmask, which provides respiratory protection by filtering inspired air through a standard Chemical-Biological-Radio-Nuclear (CBRN) filter that removes toxic agents by adsorbance and absorbance into activated charcoal in the filter [11, 12]. Mechanically ventilated patients, however, are a subgroup for whom standard CBRN gasmasks worn on the face would not provide protection from TICs, because in an environment contaminated with TICs an exposed ventilator would intake contaminated air and transmit it directly to the patient via the ventilation circuit and patient interface (regardless of whether the interface is an endotracheal tube, a tracheostomy cannula or a noninvasive facemask), without flowing through a CBRN filter along the way. A potential vulnerability to TIC exposure thus exists for patients who are ventilated on any machine that intakes atmospheric air. This would include home ventilators that intake air directly from the surrounding environment; ICU or anesthesia ventilators that operate on compressed air from a central air compressor (where the air compressor itself might be exposed to contaminated air); and ventilators that blend compressed air or oxygen (such as from an oxygen cylinder) with ambient air. In addition, for many patients who use ventilatory support only part-time—such as neuromuscular patients on nighttime facemask ventilation—wearing a CBRN gasmask while not on the ventilator might not be a realistic method for TIC protection, because the increased resistance to inspiration caused by the CBRN filter might not be tolerable for patients in chronic respiratory failure, especially at a time of psychological stress and concomitant increased respiratory rate. Although a limited number of ventilators have been custom designed to enable attachment of a CBRN filter to the ventilator’s air intake port for filtering of TICs [13], the majority of home and hospital ventilators currently in use do not have this design feature. To the best of our knowledge, no effective solutions for these ventilated patients in an event of volatile TIC dispersion have been reported in the literature. The current study was therefore aimed at developing a single standardized, universally applicable solution for TIC protection for all mechanically ventilated patients, in both the home and hospital settings, regardless of the type of ventilator being used.

The following possible solutions for ventilated patients were initially considered: attaching a CBRN filter to the air intake port of the ventilator, enclosing the patient and the ventilator in a gas-proof tent, and attaching the ventilator air intake port to a cylinder of compressed air or oxygen, which would be closed to the environment.

Preliminary experiments showed that adapting CBRN filters to cover the air intake ports of ventilators was not a feasible mass solution, because the configurations of air-intake ports differ between different ventilator models, for some of which it was impossible to attach a CBRN filter in a way that would reliably prevent contaminated air from entering. Gas-proof tents are prohibitively expensive, and erecting them is labor-intensive, making them, too, an unrealistic mass solution for all ventilated patients [14]. In addition, use of such tents is not feasible in ICUs, where rapid and unhindered access to the ventilated patient is essential at all times. Nor are such tents suitable for use in chronic-care and rehabilitation facilities, where a single member of staff may be responsible for, and therefore need repeated access to, several patients simultaneously. Cylinders of compressed gas were found to be appropriate for ICU ventilators that operate on compressed air, but not for home ventilators that draw in room air; moreover, this solution would only be effective for a limited period of time, until the contents of the cylinder ran out.

We therefore determined that the only practical, affordable and universally applicable solution to the problem of CBRN protection for ventilated patients during a TIC dispersion incident would be to attach a CBRN filter not to the air-intake port of each ventilator, but to the air-outflow port. This would ensure that even if the ventilator itself was contaminated, all air reaching the patient would be adequately filtered of contaminants; in addition, this solution would ensure that the period of time during which a ventilator could operate during suspected exposure to the toxic compounds would be practically unlimited, and that caregivers would have unhindered access to the ventilated patient at all times. Since the diameter of air-outflow ports on all ventilators conforms to a universal 22 mm standard, mass-production of a CBRN filter that would fit onto the air-outflow port of any model of ventilator in an airtight manner would be both possible and inexpensive.

The solution of attaching a CBRN filter to the air-outflow port of a ventilator, however, raised concerns that the increased resistance to airflow caused by the filter might impact adversely on the efficacy of the patient’s ventilation. For this reason, we elected to evaluate the safety, in an animal model, of adding a CBRN filter to the air-outflow tract of standard mechanical ventilators.

Methods

Animal study

A standard M-80 CBRN filter (Shalon Industries, Kiryat Gat, Israel) was customized to fit standard 22 mm diameter ventilation tubing (Fig. 1). The M-80 filter is commonly issued to both civilians and military personnel in Israel, and is in compliance with generally accepted volume and resistance standards worldwide. A healthy 30 kg Landrace pig was anesthetized with a combination of Diazepam and Ketamine, and intubated with a 7 mm endotracheal tube. The animal was then ventilated using three different models of ventilators representing typical examples of home and ICU ventilators in Israel (Synchrony, Respironics Inc, Murraysville PA; iVent, Versamed Ltd, Israel; and LTV-1000, Pulmonetic Systems, Minneapolis MN), and 17 different ventilation protocols that were designed to be representative of typical home and ICU ventilation settings (Table 1). For each ventilator model and ventilation protocol, the animal was ventilated first without, and then with, a CBRN filter added in-line to the ventilation tubing. Physiological indices (oxygen saturation, inspired CO2 [FiCO2], end tidal CO2 [etCO2], inspired oxygen [FiO2], respiratory rate [RR], and heart rate [HR]) and airflow parameters (peak inspiratory pressure [PIP], positive end expiratory pressure [PEEP], and tidal volume [TV]) were measured with and without the CBRN filter in place. Protocols 1 through 3 used volume cycled ventilation, protocols 4 through 6 used time-cycled pressure-limited ventilation, and protocols 7 through 12 used flow-cycled (pressure support) ventilation, as typically supplied by bilevel ventilators. For ventilation protocols 1 through 12 the CBRN filter was positioned between the ventilator and the animal’s respiratory circuit (Fig. 1). This was assumed to be the ideal positioning for the CBRN filter, as it achieves CBRN filtration of all inspiratory gas flow, without impacting on expiratory gas flow at all. However, so as to evaluate the potential physiological impact of the CBRN filter when exposed to expiratory airflow, in protocols 13 through 16, using time-cycled pressure-limited ventilation, the CBRN filter was positioned between the animal’s respiratory circuit and endotracheal tube, such that both inspiratory and expiratory gas flow passed through the filter (Fig. 2). Between changes in protocols 1 through 16, and after addition or removal of the CBRN filter, the animal was allowed to stabilize for 3–5 min before physiological and airflow measurements were made. So as to evaluate the potential impact of a longer period of ventilation through an CBRN filter, in protocol number 17 volume-cycled ventilation through a CBRN filter was maintained uninterrupted for 30 min before physiological measurements were recorded.

Fig. 1
figure 1

CBRN filter attached to the air-outflow port of a home ventilator

Table 1 Ventilation protocols used
Fig. 2
figure 2

CBRN Filter positioned between an endotracheal tube and the respiratory circuit of a home ventilator

Throughout the trial the animal’s state of anesthesia was maintained at a level that prevented the animal from actively opposing mechanical ventilation, and all physiological measurements were made with the animal stable on room air without supplemental oxygen. The trial was approved by the Animal Care Committee of Rambam Medical Center and was conducted at the Animal Research Laboratory of the Technion Medical School/Rambam Medical Center in Haifa, Israel. Animal care and handling were in accordance with the National Institute of Health guidelines for ethical animal research.

In vitro study

Since the impact of a CBRN filter on a ventilator’s trigger/sensitivity function (i.e., the negative pressure or flow that a patient has to develop spontaneously so as to trigger a mechanical breath from the ventilator) cannot be evaluated on an anesthetized animal that is not breathing spontaneously, this parameter was studied in vitro. First, the resistance versus flow characteristics of a standard-issue CBRN filter were measured empirically using a Windjammer 5.1 inch blower (Amatek Inc. Kent, OH) and an IMT PF300 Flow Analyzer (IMT Medical, Buchs, Switzerland). Then, a Michigan test lung and a Synchrony ventilator set to deliver PEEP at levels of 5, 7 and 9 cm H2O were used to evaluate the impact of the CBRN filter on the ventilator’s trigger/sensitivity function. Spontaneous inspiratory effort was simulated by randomly generating brief episodes of negative pressure within the test lung at each of the PEEP values. The negative-pressure waveform that propagated into the ventilation tubing as a result was recorded by two pressure-flow sensors (Y Flow Transducer, Versamed Ltd., Israel) connected in-line between the test lung and the ventilator. The pressure gradient between the sensors was measured first without the CBRN filter in situ, and thereafter with the CBRN filter inserted between the two sensors. The pressure gradient between the pressure-flow sensors was measured using SV Monitor software. The internal volume of the CBRN filter was measured using a water immersion and extraction technique.

Results

Animal study

The animal was ventilated for a total of six and a half hours, for approximately half of which the CBRN filter was in situ. Table 2 shows the airflow parameters measured for each ventilation protocol. On average, positioning the CBRN filter in the ventilation circuit resulted in a 16 ml (5 %) decrease in tidal volume (range 0–50 ml), a 1.7 cm H2O (10 %) decrease in peak inspiratory pressure (range 1–3 cm H2O), and a 0.1 cm H2O (3 %) decrease in PEEP across the filter (range 0–1 cm H2O). The way that the ventilators responded to the changes caused by addition of the CBRN filter to the ventilation circuit depended on the design of the ventilator’s pressure sensor. The Synchrony machine has an internal pressure/flow sensor located within the ventilator. For this reason, in protocols in which the CBRN filter was positioned on the gas outflow port of the Synchrony ventilator (protocols 7–12), the filter was “downstream” from the ventilator’s pressure/flow sensor. In these instances, the ventilator did not sense or compensate for the pressure/volume drop-off across the filter after the filter was added. The animal consequently received somewhat less ventilation with the filter in situ than without. In contrast, the pressure/flow sensor of iVent and LTV machines is located at the far end of the ventilation tubing, close to the patient. For this reason, the CBRN filter positioned on the gas outflow port of these ventilators (protocols 1–6 and 13–17) was “upstream” from the ventilator’s pressure/flow sensor, enabling the ventilator to sense and compensate for the pressure/volume drop-off caused by the addition of the filter. The animal thus continued to receive the same degree of ventilation even after the addition of the filter.

Table 2 Change in air-flow parameters caused by addition of CBRN filter to circuit

Addition of a filter between the ventilator and the respiratory circuit (protocols 1–12 and protocol 17) did not cause any detectable deterioration in the animal’s respiratory physiology (saturation, etCO2, RR and HR remained constant), and there was no evidence of rebreathing (no change in FiO2 or FiCO2). However, when the CBRN filter was positioned between the respiratory circuit and the endotracheal tube (protocols 13 through 16), evidence of rebreathing emerged within 1 min: FiCO2 increased to 1 %, FiO2 decreased to 17 %, and oxygen saturation decreased to less than 90 %.

In vitro study

The resistance versus flow characteristics of the CBRN filter as measured empirically is presented in Fig. 3. Table 3 shows the impact of the CBRN filter on the trigger/sensitivity function of a Synchrony ventilator. The table shows the negative-pressure (“trigger”) gradient generated by the Michigan test lung, as measured by two pressure-flow sensors positioned between the test lung and the ventilator. The negative pressure inspiratory flow generated by each of the simulated inspiratory efforts (i.e. prior to the onset of positive pressure airflow from the ventilator) ranged from 1 to 5 L/min. Without the filter in situ, the intrinsic resistance of the sensors resulted in a mean pressure gradient of 0.14 cm H2O between them. After insertion of the filter between the sensors, generation of negative pressure spikes in the test lung showed a mean pressure gradient of 0.59 cm H2O (range 0.19–1.39 cm H2O) across the filter. Ignoring the intrinsic resistance of the sensors, the filter generated a mean pressure gradient of 0.45 cm H2O, a one-third drop-off in trigger pressure. The internal volume of the CBRN filter, representing the dead space added to the respiratory airflow pathway by addition of the filter, was measured as being 140 ml.

Fig. 3
figure 3

Resistance vs flow of a standard-issue CBRN filter

Table 3 Trigger gradient across the CBRN filter (cm H2O)

Discussion

Manufacturing and chemical plants may pose environmental hazards to nearby residents, due to either accidents or malpractice [110]. In addition, in the current era biological and chemical TICs have been used in warfare and acts of terrorism. Ventilated patients in institutions and ambulatory settings may thus be affected by TICs. Nevertheless, current protective protocols do not provide an adequate solution for this population. In this study, the safety of a standard CBRN filter added in-line to the ventilation circuit of an animal model was tested. We show that the addition of a CBRN filter is safe, well tolerated physiologically, and has minimal impact on airflow parameters. It is suggested that this solution may be a useful method for protecting ventilated patients in all settings from TICs.

We are aware of several limitations to this study. Technical constraints prevented us from evaluating the loss of tidal volume across the CBRN filter under conditions of volume-cycled ventilation. Changes in this parameter, however, are unlikely to differ from those found under conditions of time-cycled and flow-cycled ventilation. In addition, inspiratory time was not measured, as this parameter was fixed for all the ventilation protocols used in the trial and thus would not change with the addition of a CBRN filter to the ventilation circuit.

Application of this technique on a mass scale may require several precautions in order to minimize the risk of harm being caused to the patient by the filter. It is important to ensure that the CBRN filter is positioned directly on the air outflow port of the ventilator, and not on the patient’s tracheostomy or endotracheal tube, as this will lead to dangerous rebreathing of exhaled CO2, most likely due to accumulation of CO2 in the dead space of the CBRN filter. Although off-gassing (i.e. second-hand exposure to toxins in air exhaled from a contaminated patient) is a concern for caregivers treating patients exposed to nerve agents, and positioning the CBRN filter directly on the patient’s endotracheal tube might provide some protection against this, for most TICs this is not a real concern, and the danger to the patient of hypercapnia from rebreathing far outweighs the theoretical benefit of minimizing off-gassing by placing the CBRN filter directly on the endotracheal tube.

For ventilators with an externally located pressure/flow sensor, such as most volume-cycled and time-cycled ventilators, no readjustment of tidal volume or PIP should be necessary after addition of a CBRN filter to the ventilation circuit, provided the CBRN filter is correctly positioned on the ventilator outflow port, “upstream” from the pressure/flow sensor. However, for ventilators with an internally located pressure/flow sensor, such as bi-level ventilators, it may be advisable to increase the inspiratory positive airway pressure (IPAP) by 2 cm H2O so as to compensate for potential pressure drop-off after addition of a CBRN filter. Although some patients on home ventilation will be able to perform such an adjustment competently and safely, and will remember to return the ventilator settings to their baseline after removal of the CBRN filter, the potential for error on the part of less aware home ventilated patients is real, particularly during a time of high stress and anxiety. For this reason, public health authorities should consider whether or not adequate education of this population is feasible before issuing a general recommendation to change the parameters of bi-level ventilators when attaching a CBRN filter. It may be more prudent to recommend that healthcare workers make such changes, wherever possible, while advising patients not to adjust their ventilator settings themselves.

Patients breathing spontaneously should be advised that the addition of the filter may cause a sensation of increased effort of breathing, due to the increase in resistance to trigger flow caused by the CBRN filter. This may necessitate increasing the mandatory number of breaths delivered by the ventilator per minute as a means of compensation. In addition, the sensation of increased resistance to initiating a breath that is caused by the addition of a CBRN filter may be mitigated in part by increasing the patient’s PEEP setting or adjusting the sensitivity of the ventilator’s trigger function.

Both public health authorities and patients should be informed as to the recommended duration of use of a CBRN filter once it has been unsealed and inserted into the patient’s ventilation circuit. When exposed to the TIC ambient air concentrations that are typically encountered in industrial accidents or acts of terror, CBRN filters provide adequate protection for a minimum of 8 h, at all clinically used ventilation minute volumes. There is therefore no need to replace the filter prior to that, and it is highly unlikely that a single TIC exposure incident would last longer than that timeframe. However, in the event of prolonged exposure beyond 8 h, a new filter should be inserted as soon as is feasible thereafter. If a new filter has been opened but not exposed to active airflow, it can be resealed and stored for up to 12 months without loss of efficacy. Although addition of a CBRN filter to a patient’s ventilation circuit protects the patient from inhaling TICs, the patient remains exposed to these agents through his mucous membranes. For this reason, patients being ventilated via a tracheostomy should wear a standard gasmask to protect the mucous membranes of their eyes, nose and mouth, in addition to adding a CBRN filter to their ventilation tubing. This may not be feasible, however, for patients being ventilated via an endotracheal tube or facemask.

Another concern is that of a possible exothermic reaction between the charcoal of the CBRN filter and enriched oxygen, which is commonly used by ventilated patients. This was tested separately in a study that showed no such reaction, and that oxygen flow through a CBRN filter is safe [15].

Conclusions

In-line addition of a CBRN filter to the ventilation circuit represents a feasible strategy for affording inhalational protection from toxic inhalational compounds to the general population of home and ICU ventilated patients. Further clinical studies are necessary to confirm these findings.

Abbreviations

CBRN:

Chemical-Biological-Radio-Nuclear

TICs:

toxic industrial compounds

PEEP:

positive end expiratory pressure

PIP:

peak inspiratory pressure

TV:

tidal volume

FiCO2 :

inspired CO2

etCO2 :

end tidal CO2

FiO2 :

inspired oxygen

RR:

respiratory rate

HR:

heart rate

IPAP:

inspiratory positive airway pressure

Ti:

inspiratory time

References

  1. Hughart JL, Bashor MM. Industrial chemicals and terrorism: human health threat analysis, mitigation and prevention. Agency for toxic substances and disease registry, U.S. Public Health Service. http://www.mapcruzin.com/scruztri/docs/cep1118992.htm. Accessed 26 Sept 2015.

  2. Anderson AR, Wu J, Centers for Disease Control and Prevention (CDC). Top five industries resulting in injuries from acute chemical incidents—Hazardous Substance Emergency Events Surveillance, nine states, 1999–2008. MMWR Surveill Summ. 2015;64(Suppl 2):47–53.

    PubMed  Google Scholar 

  3. Williams BL, Magsumbol MS. Emergency preparedness among people living near US army chemical weapons sites after September 11, 2001. Am J Public Health. 2007;97(9):1601–6.

    Article  PubMed Central  PubMed  Google Scholar 

  4. Burklow TR, Yu CE, Madsen JM. Industrial chemicals: terrorist weapons of opportunity. Pediatr Ann. 2003;32:230–4.

    Article  PubMed  Google Scholar 

  5. Bennett RL. Chemical or biological terrorist attacks: an analysis of the preparedness of hospitals for managing victims affected by chemical or biological weapons of mass destruction. Int J Environ Res Public Health. 2006;3:67–75.

    Article  PubMed Central  PubMed  Google Scholar 

  6. Koplan JP, Holloway BR, Jackson RJ, Hughes JM, Rosenstock L, Baker EL. Biological and chemical terrorism: strategic plan for preparedness and response. Recommendations of the CDC strategic planning workgroup. MMWR. 2000;49(RR-4):1–14.

    Google Scholar 

  7. Moore DH, Alexander SM. Emergency response to a chemical warfare agent incident: domestic preparedness, first response and public health considerations. In: Somani SM, Romano JK, editors. Chemical warfare agents: toxicity at low levels. Boca Raton: CRC Press; 2001.

    Google Scholar 

  8. Rogers GO, Sorensen JH, Watson AP. Protecting civilian populations during chemical agent emergencies. In: Somani SM, editor. Chemical warfare agents. New York: Academic Press; 1992.

    Google Scholar 

  9. Sidell FR. Chemical agent terrorism. Ann Emerg Med. 1996;28:223–4.

    CAS  PubMed  Google Scholar 

  10. Rosman Y, Eisenkraft A, Milk N, Shiyovich A, Ophir N, Shrot S, Kreiss Y, Kassirer M. Lessons learned from the Syrian sarin attack: evaluation of a clinical syndrome through social media. Ann Intern Med. 2014;160(9):644–8.

    Article  PubMed  Google Scholar 

  11. Rimpel LY, Boehm DE, O’Hern MR, Dashiell TR, Tracy MF. Chapter17: Chemical Defense Equipment. In: Medical aspects of chemical warfare. In: Tuorinsky SD, editor. Textbook of military medicine. Office of the Surgeon General, US Army, Borden Institute, Walter Reed Army Medical Center, Washington DC. 2008. p. 559–92.

  12. Martyny J, Glazer CG, Newman LS. Respiratory Protection. N Engl J Med. 2002;37:824–30.

    Article  Google Scholar 

  13. Roberts MJ, Bell GT, Wong LS. The CompPAC and PortaPAC portable ventilators bench tests and field experience. J R Army Med Corps. 1999;145(2):73–7.

    Article  CAS  PubMed  Google Scholar 

  14. Health Care Resources. Department of Health Information, Ministry of Health, Israel [Hebrew]. 2002.

  15. Krivoy A, Luria S, Gafnit Z, Sukhotnikovt S, Yehezkelli Y. Safety of biological/chemical respiratory protection filter for patients in need of oxygen supplementation. Mil Med. 2005;170(12):1029–31.

    Article  PubMed  Google Scholar 

Download references

Authors’ contributions

EB conceived the study. EB, MS, YB and AE participated in the design of the study. All authors participated in the study and data collection. EB and SO performed the statistical analysis. EB and AE drafted the manuscript. All authors read and approved the final manuscript.

Competing interests

MS was employed by Versamed Ltd, Israel, at the time of this study. EB, SO, YB and AE declare no funding source for this research.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Arik Eisenkraft.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Be’eri, E., Owen, S., Shachar, M. et al. Impact of adding a filter for protection from toxic inhalational compounds to the ventilation circuit of mechanically ventilated patients. Disaster and Mil Med 2, 3 (2016). https://doi.org/10.1186/s40696-016-0015-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40696-016-0015-6

Keywords